Interval Trigger

Trigger when the times difference between adjacent rising or falling edges meets
the time limit (<=, >=, [--,--], --][--).
To set interval trigger:
1. Press the Setup button on the front panel to enter the TRIGGER system function
menu.
2. Press the Type softkey; then use the Universal Knob to select Interval and push
down the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1 or CH2 as the
trigger source.
4. Press the Slope softkey to select rising or falling edge.
5. Press the Limit Range softkey; turn the Universal Knob to select desired condition.
<= (less than a time value): trigger when the positive or negative pulse time of the
input signal is lower than the specified time value.
>= (greater than a time value): trigger when the positive or negative pulse time of
the input signal is greater than the specified time value.
[--,--] (within a range of time value): trigger when the positive or negative pulse
time of the input signal is greater than the specified lower limit of time and lower
than the specified upper limit of time.
--][-- (outside a range of time value): trigger when the positive or negative pulse
time of the input signal is greater than the specified upper limit of time or lower
than the specified lower limit of time value.
6. Press the Time Setting softkey (<=, >=, [--,--],--][--), turn the Universal Knob to
select the desired value.