Figure 3-11 V - I characteristic curve of Von Latch

Set Trigger

Trigger operation is mainly applied to the Dynamic, List and Program test
functions. The load supports three modes: Manual, Ext and Bus.
� Manual
When the trigger source is set to Manual mode, press the trig tab in
dynamic, List or Program mode will generate a trigger signal.
� Bus
When the trigger source is set to Bus mode, the load will execute a
trigger operation when it receives the remote command (such as
*TRG) from USB, LAN, RS232 and GPIB interface.
The Ext monitor terminals on the rear panel of the load can receive external
trigger signals. When the trigger source is set to be Ext mode and receives a
pulse (0-5V TTL) which has a falling edge, the load will trigger.

Set Average

In "Config" tab, when user select "Aver" can set the average point number of the
read-back current and voltage in the load. The range is 6-14. The greater of
the average point number, the slower of the read-back current and voltage in
the load.
76