Lin Triggering And Serial Decode; Setup For Lin Signals - Teledyne T3DSO1000A Series User Manual

Digital oscilloscope
Table of Contents

Advertisement

LIN Triggering and Serial Decode

LIN Trigger and decode steps: "Setup for LIN Signals", "LIN Triggering" and
"LIN Decode" to trigger and decode the signals.

Setup for LIN Signals

There are two steps for setting the LIN signal, connecting the signal to oscilloscope,
specifying the parameters of each input signal.
1.
Press the Decode key to enter the DECODE function menu.
2.
Press the Decode softkey and select the desired decoder (Serial 1 or Serial 2).
3.
Press Protocol softkey and then select LIN by turning Universal Knob.
4.
Press Signal softkey to enter the SIGNAL menu as Figure 54 shows.
5.
Press Source softkey to select the channel that is connected to the LIN signal.
6.
Press the Threshold softkey and set the LIN signal's threshold voltage level using the
Universal Knob. The threshold voltage level is for decoding, and it will be used as
the trigger voltage level when setting the trigger type to serial.
7.
Press
softkey to return to the previous menu.
8.
Press the Configure softkey to enter the BUS CONFIG menu.
9.
Press Baud softkey to set baud rate.
The baud rate can be set as a predefined value.
If the desired baud rate is not listed, select custom option, press the Custom
and turn the Universal Knob to set the desired baud rate.
Figure 54 LIN SIGNAL Menu
User Manual 109