WARNING
To avoid electric shock during the use of a probe, please make sure that the
insulated wire of the probe is in good condition and do not touch the metallic
part of the probe when the probe is connected to a high voltage source
Probe Compensation
When the probes are used for the first time, you should compensate the probes to match
the input channels of the oscilloscope. Non-compensated or poorly compensated probes
may cause measurement inaccuracy or error. The probe compensation procedures are as
follows.
1. Set the switch to 10X on the probe.
2. Perform steps 1, 2, 3 and 4 of "Function Inspection" in the previous section.
3. Check the waveforms displayed and compare them with the following:
Over
Compensated
4. Use a nonmetallic screwdriver to adjust the low-frequency compensation adjustment
hole on the probe until the waveform displayed is as the "Perfectly compensated"
waveform in the figure above.
Perfectly
Compensated
Under
Compensated
User Manual 9