Trigger Coupling - Teledyne T3DSO1000A Series User Manual

Digital oscilloscope
Table of Contents

Advertisement

Trigger Coupling

Press the Setup button on the front panel to enter the TRIGGER function menu, and then
press the Coupling softkey and turn the Universal Knob or press the Coupling softkey
continually to select the desired coupling mode.
The oscilloscope provides 4 kinds of trigger coupling modes:
DC: allow DC and AC components into the trigger path.
AC: block all the DC components and attenuate signals lower than 5.8 Hz. Use AC
coupling to get a stable edge trigger when your waveform has a large DC offset.
● LF Reject: block the DC components and reject the low frequency components lower
than 2.08MHz. Low frequency reject removes any unwanted low frequency
components from a trigger waveform, such as power line frequencies, etc., that can
interfere with proper triggering. Use LF Reject coupling to get a stable edge trigger
when your waveform has low frequency noise.
● HF Reject: reject the high frequency components higher than1.27MHz)
Note: trigger coupling has nothing to do with the channel coupling.
User Manual 57

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3dso1000 seriesT3dso1102T3dso1204T3dso1104

Table of Contents