To Specify Channel Coupling; To Specify Bandwidth Limit - Teledyne T3DSO1000A Series User Manual

Digital oscilloscope
Table of Contents

Advertisement

To Specify Channel Coupling

Set the coupling mode to filter out the undesired signals. For example, the signal under
test is a square waveform with DC offset.
When the coupling mode set to DC: the DC and AC components of the signal under
test can both pass into the scope measurement channel.
When the coupling mode set to AC: the DC components of the signal under test are
blocked.
When the coupling mode set to GND: the DC and AC components of the signal under
test are both blocked.
Press the CH1 button on the front panel; then press the Coupling softkey and turn the
Universal Knob to select the desired coupling mode. The default setup is DC.
The current coupling mode is displayed in the channel label at the right side of the screen.
You can also press the Coupling softkey continuously to switch the coupling mode.

To Specify Bandwidth Limit

Set the bandwidth limit to reduce display noise. For example, the signal under test is a
pulse with high frequency oscillation.
When the bandwidth limit set to Full, the high frequency components of the signal
under test can pass into the scope measurement channel.
When the bandwidth limit set to 20M, the high frequency components that exceed 20
MHz are attenuated.
Press the CH1 button on the front panel; then press the BW Limit softkey to select Full or
20M. The default setup is Full. When bandwidth limit is enabled, the character B will be
displayed in the channel label at the right side of the screen.
T3DSO1000 series has full BW at all V/div settings including 500uV/div to 2mV/div.
28 User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3dso1000 seriesT3dso1102T3dso1204T3dso1104

Table of Contents