Average
In this mode, the oscilloscope averages the waveforms from multiple acquisitions to
reduce the random noise of the input signal and improve the vertical resolution. The
greater the number of averages, the lower the noise will be, and the higher the vertical
resolution will be, but the slower the response of the displayed waveform to waveform
changes.
The available range of averages is from 4 to 1024 and the default is 16. When Average
mode is selected, press Averages and turn the universal knob or press the softkey
continually to set the desired average time.
Figure 21 With Random Noise, Normal Mode