Replacement of the Caps for TGC Sliders and/or Rotation Digipots

NOTE:
Please observe that Caps for TGC Sliders and Rotation Digipots depend on installed Console (RTU).
The Console version is shown in the System Info page (see:
8-14-1
Manpower
One person, 5 minutes
8-14-2
Tools
none (poss. small-sized slotted screwdriver or tweezers)
8-14-3
Caps for TGC Sliders and/or Rotation Digipots - Replacement Procedure
1.) Remove the caps for Slider-potentiometer TGC (and/or Rotation Digipots).
NOTE:
Don't loose integrated metal spring in each Rotation Digipot cap.
2.) Mount the caps for Slider-potentiometer TGC (and/or Rotation Digipots).
8-22
Section 8-14 - Replacement of the Caps for TGC Sliders and/or Rotation Digipots
7
DRAFT (A
EVISION
Figure 8-24 remove caps
23, 2012)
UGUST
Figure 7-1 on page
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
7-3).