KTD102576, R
IRECTION
4-4-11
Patient Archive (Image Management)
The Voluson E8 / Voluson E6 provides an Patient/Image Management System that allows fast and
extremely easy patient, exam and image management.
Current Patient:
The entered patient data will be used in calculations, patient worksheets, DICOM settings and is
displayed on the screen to identify images.
Archive:
The patient archive database is used for searching a particular exam and/or patient.
Via the DATA TRANSFER button, it is possible to send images over the DICOM network, print
exams/images, export exams/images, import exams/images, etc.
Image History:
Image History gives you access to all the US pictures and exams of a particular patient.
Exam Review:
Exam Review allows you to view all exams of a particular patient. Additionally it is possible to view
image properties, input comments and voice annotations, etc.
NOTE:
For detailed description of image management functionality and its topics refer to the Voluson E8 /
Voluson E6 Basic User Manual.
NOTE:
Images can be backed up and restored by means of the Image Archive function in the System Setup.
Operation see:
4-30
7
DRAFT (A
EVISION
Figure 4-19 Patient Archive - ARCHIVE
Section 4-5-6 "Archiving Images" on page
Section 4-4 - Functional Checks
23, 2012)
UGUST
4-42.
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
E6