KTD102576, R
IRECTION
4-5-6
Archiving Images
It is recommended to Save Image Archive
CAUTION
System Configuration
!! CAUTION:
4-5-6-1
Save Image Archive
NOTE:
A backup of the Image Archive always contains the Patient Archive (database containing patient
demographic data and measurements) + images of the selected exams.
1.) Press the UTILITIES key on the control panel.
2.) In the "Utilities" menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP and then click the
4.) Click the SAVE button.
4-42
Section 4-5 - Backup and Restore Database, Preset Configurations and Images
7
DRAFT (A
EVISION
(Section 4-5-3 on page
Figure 4-35 System Setup - Backup - IMAGE ARCHIVE page
Figure 4-36 Image Archive Save - preparations
23, 2012)
UGUST
(Section 4-5-6-1 on page
4-37) once a week.
V
E8 / V
OLUSON
OLUSON
S
ERVICE
4-42) and Backup the Full
IMAGE ARCHIVE
tab.
[Load] button
[Save] button
[Advanced] button
choose archiving
span time
check mark if desired
click [Next] button