Connecting the 19" LCD Secondary "Patient" Monitor

The 19" LCD Secondary "Patient" Monitor MUST NEVER be connected to the Voluson E8 /
CAUTION
Voluson E6 ultrasound systems mains supply directly!
!! CAUTION:
Always connect it to the supplied Isolation Transformer!
The Secondary Monitor is the only item to be connected to the Transformer.
The 19" LCD Secondary Monitor is NOT intended for diagnostic use.
NOTICE
It is an additional device used to allow the patient to watch the proceedings.
!! NOTICE:
Take your time to think about the best position of the monitor in your facilities.
Patients should be able to view the monitor easily and without having to bend or turn around.
DO NOT connect the 19" LCD Secondary Monitor to the Voluson E8 / Voluson E6 via USB cable.
NOTICE
Use the supplied VGA cable.
!! NOTICE:
NOTICE
For more detailed description, how to replace/mount the 19" LCD Secondary "Patient" Monitor, refer to:
Chapter 8 - Replacing optional Peripherals / How to mount Peripherals at a later date, on page
!! NOTICE:
1.) Power OFF/Shutdown the system as described in:
2.) Connect the Secondary Monitor according to connection scheme (see:
NOTE:
Connection of the 19" LCD Secondary "Patient" Monitor is always the same
(no differences between PC-Motherboard and/or BT-version of the system).