GE Voluson E8 Service Manual page 316

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
System detected severe error. Some components like Touch Panel server may not
be registered. Please register Touch Panel server and restart.
Action: If after restart not ok then call technical support
System must be restarted. Reason: Corrupt Factory Settings
System must be restarted.\nReason: Corrupt Factory Settings. Choose ""Ok"" to use
Safety-Copy after Restart. Choose ""Cancel"" and load Factory Settings from
Backup after restart.
The "Current Password" does not match the actual password.
The "New Password" does not match with "Retype new password".
The Database UserPrograms Corrupted. Action: Restart the system
The Date format not stored properly. Action: close registry, restart, try again
The device can now be safely removed from the system.
The entered password does not match the actual password.
The Handle Unregistered. Action: close registry, restart, try again
The Registry not closed. Action: close registry, restart, try again
The System detected a problem with your harddisk.
Please do not reboot or shut down the machine!
Please contact your service department as soon as possible.
There is no Printer selected!
Thickness mismatch xx - GIP xx.
Action: restart 3D (go to 2D), restart the system
throw HRS_Exception(err_msg);
throw HRS_Exception(error.ErrorMessage());
throw HRS_Exception(error.what());
throw HRS_Exception(str);
Tx-Power to HW : Unkown Systemtype.
Action: restart the system, If after restart not ok then call technical support
UI_BBC_ArchiveWnd : incomplete implementation for....
UI_BBC_ArchiveWnd::vSet() has an wrong ImageType. Action: restart the system
UI_BBC_ArchiveWnd::vSet() will change from eB_Wnd to wrong ImageType.
Action: restart the system
UI_BBC_ArchiveWnd::vSet() will change from eBBC_Wnd to wrong ImageType.
Action: restart the system
UI_BBC_Wnd::vSet() has an wrong ImageType.
Action: restart the system
7-44
7
DRAFT (A
EVISION
Error Messages
Section 7-9 - Error Messages
23, 2012)
UGUST
restart the system, call technical support
restart the system
restart the system
enter correct password
retype both passwords
restart the system
close registry, restart the system and try again
press ok and remove the USB volume
enter correct password
close registry, restart the system and try again
restart the system
if possible,
• save "Full Backup" (see:
"DVD/CD", "Network", or "Other drive" (including images)
• if not currently done, save "Image Archive"
(see:
Section 4-5-6-1 on page
• call technical support
select a printer
restart 3D (go to 2D); restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
Actions
Section 4-5-3 on page
4-37) to
4-42),
E6

Advertisement

Table of Contents
loading

Table of Contents