GE Voluson E8 Service Manual page 351

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
8-17-4-4
Preparing the Isolation Transformer (cont'd)
4.) Push the fuse holder back into position (correct voltage should be shown in upper line; e.g, 230V,
see:
NOTE:
The adjusted voltage is shown in the viewing window (see:
5.) The output voltage must also be changed by the turn switch, which is located on the right side.
For that purpose use an appropriate screwdriver and make sure the transformer is switched off.
-
-
voltage
(e.g. 230V)
is shown in
this window
6.) Assure that the connected loads can be operated with the chosen voltage.
7
DRAFT (A
EVISION
Figure 8-43
below) and then close fuse protection.
Fuse holder
(e.g. 230V)
Figure 8-43 change fuses of isolation transformer
for an Input voltage range of 100-130V set position A
for an Input voltage range of 220-240V set position B
Figure 8-44 output voltage setting
Chapter 8 - Replacement Procedures
23, 2012)
UGUST
Figure 8-44
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
below).
voltage output
selector
E6
8-33