GE Voluson E8 Service Manual page 350

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
8-17-4-4
Preparing the Isolation Transformer
Isolation Transformer
Power Cord set
Figure 8-40 Isolation Transformer kit for 19" LCD Secondary Monitor
1.) Before using the Isolation Transformer you must check the input voltage settings to meet the ratings
of the line power available in your location or country.
NOTICE
The wrong fuses and position of the voltage output selector may cause major damage on connected
peripherals.
!! NOTICE:
2.) For changing the input voltage, open the fuse protection at the power inlet block with a small
screwdriver and remove the fuse holder.
voltage
(e.g. 230V)
is shown in
this window
3.) By turning the fuse holder you can select 115V or 230V.
Consider that changing the input voltage also requires to change the fuses!
8-32
Section 8-17 - Replacing optional Peripherals / How to mount Peripherals at a later date
7
DRAFT (A
EVISION
Figure 8-41 changing input voltage
Figure 8-42 input voltage setting
23, 2012)
UGUST
voltage output
selector
115V (100-130V) -> 4AT
230V (220-240V) -> 2AT