Connecting The Vga Image (Video) Resizer - GE Voluson E8 Service Manual

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
3-5-7

Connecting the VGA Image (Video) Resizer

NOTICE
For more detailed description, how to replace/mount the VGA Image Resizer, refer to:
Chapter 8 - Replacing optional Peripherals / How to mount Peripherals at a later date, on page
!! NOTICE:
NOTICE
The VGA Image (Video) Resizer is required whenever the used Secondary "Patient" Monitor has a
different screen resolution than the Voluson E8 / Voluson E6 system!
!! NOTICE:
Console back
S - Video
DVI - Out
Out
Out
In
NOTE:
After physical connection of the VGA Image Resizer to the Voluson system, follow the procedure
described in
7
DRAFT (A
EVISION
DVD Writer
text
Lift is only used for the lifter
(230V - can not be changed)
Figure 3-18 Connection Scheme - VGA Image Resizer
Section 3-5-7-1 "Adjustment of the VGA Image Resizer Settings" on page
Chapter 3 - Setup Instructions
23, 2012)
UGUST
Power Supply
A Secondary "Patient" Monitor MUST NEVER be connected to
the Voluson E6 / Voluson E8 systems mains supply directly!
Always connect it to an appropriate Isolation Transformer
(see:
Chapter 9 - Optional Equipment, on page
Connect the power supply of the VGA Image Resizer to the
power cable plug behind the door on the right side cover.
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
8-25.
VGA Image
Resizer box
to Secondary
"Patient" Monitor
CAUTION
!! CAUTION:
9-33)
3-24.
E6
3-23

Advertisement