GE Voluson E8 Service Manual page 314

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
RT_4DTissueFilterBlock:: Storage Error, no dynamic memory for filter operations
available!!
RT_4DTissueFilterBlock:: storage needed for one filtered volume differs from
available Unitsize within replay buffer!!
RT_4DTissueFilterBlock::bDllLineFilter call failed!!
RT_4DTissueFilterBlock::bDllLineFrameFilter call failed!!
RT_4DTissueFilterBlock::DMA Block size and calculated frame size differs!!
RT_4DTissueFilterBlock::execute replay buffer write address is NULL!!
Action: restart the system, If after restart not ok then call technical support
RT_4DTissueFilterBlock::vCheckCahngedInports Replay buffer not cleared before
VolGeoChange!!
RT_ColorFlowFilterBlock Constructor: Memory allocation failed
RT_ColorFlowFilterBlock::vCheckIQDataSizeAndUpdateTables: Memory allocation
failed. Action: restart the system, If after restart not ok then call technical support
RT_ColorFlowFilterBlock::vDebugDrawIQDataCurve: Memory allocation failed.
Action: restart the system, If after restart not ok then call technical support
RT_ConnectionMngr::vAssembleRTSet: no RT_TissueFilterBlock found for
TISSUE3D Blocks
RT_ECG_Block::bStart ECGInterface failed!!
RT_MColorFilterBlock::Memory allocation failed.
Action: restart the system, If after restart not ok then call technical support
RT_MColorFilterBlock::vCheckIQDataSizeAndUpdateBuffer: Memory allocation
failed
RT_MColorFilterBlock::vDebugDrawIQDataCurve: Memory allocation failed
RT_MotionMBlock::execute caused exception:: Addr from DMA= xx,
ReplayBuffAddr= xx, (Line+Header)Size= xx, blockLength= xx, value i= xx!! Action:
restart the system, If after restart not ok then call technical support
RT_MotionMCTestBlock::execute: Error, length of DMA block 1[xx] is not aligned to
line size!!
RT_MotionMCTestBlock::execute: Error, length of DMA block 2[xx] is not aligned to
line size!!
RT_PWorCW_Block::replay buffer size smaller than expected
RT_PWorCW_Block::execute caused exception:: Addr from DMA=xx,
ReplayBuffAddr=xx, (Line+Header)Size=xx, blockLength=xx, line index=xx!!
RT_PWorCW_Block::execute: Error, length of DMA block 1[xx] is not aligned to line
size!! Action: restart the system, If after restart not ok then call technical support
RT_PWorCWFilterBlock::execute : replay buffer size smaller than expected.
Action: restart the system, If after restart not ok then call technical support
RT_PWorCWFilterBlock::execute caused exception:: Addr from DMA=xx,
ReplayBuffAddr=xx, (Line+Header)Size=xx, blockLength=xx, line index=xx!!
Action: restart the system, If after restart not ok then call technical support
7-42
7
DRAFT (A
EVISION
Error Messages
Section 7-9 - Error Messages
23, 2012)
UGUST
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
Actions
E6

Advertisement

Table of Contents
loading

Table of Contents