IRECTION
10-7-7
Isolated Patient Lead (Source) Leakage–Lead to Lead
When using the Dale 600, switch the meter's function selector to the LEAD-LEAD position.
Select and test each of the ECG lead positions (except ALL) on the LEAD selector, testing each to the
power and ground condition combinations found in:
Table 10-10 on page
Table 10-11 on page
Record the highest leakage current measured.
10-7-7-1
Lead to Lead Leakage Test Record
Table 10-18
Measurements should be recorded from each lead combination under each set of test conditions
specified in:
Table 10-10 on page
Table 10-11 on page
Record all data on the EQC inspection certificate.
Table 10-18 Typical Data Format for Recording Patient Lead to Lead Leakage
System Under Test:
System Power
Mains on
RF on
NOTE:
Values in italics font are given as examples only.