Electrical Safety

1-3-4-1

Safe Practices

To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
The system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. If an extension cord is used with the system, make sure that the total
current rating of the system does not exceed the extension cord rating.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
CONNECTING A VOLUSON E8 / VOLUSON E6 SCANNER TO THE WRONG VOLTAGE
WARNING
WARNING
LEVEL WILL MOST LIKELY DESTROY THE SCANNER.
!! WARNING:
1-3-4-2

Probes

All the probes for the Voluson E8 / Voluson E6 are designed and manufactured to provide trouble-free,
reliable service. To ensure this, correct handling of probes is important and the following points should
be noted:
Do not drop a probe or strike it against a hard surface, as this may damage the transducer elements,
acoustic lens, or housing.
Inspect the probe prior to each use for damage or degradation to the Housing, Cable strain relief,
Lens and Seal.
Do not use a cracked or damaged probe. In this event, call GE service representative immediately
to obtain a replacement.
Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled,
or to be caught in the system's wheels.
Never immerse the probe connector or adapter into any liquid.
NOTE:
For detailed information on handling probes, refer to the Voluson E8 / Voluson E6 Basic User Manual
and the care card supplied with the probe.
1-3-5

Auxiliary Devices Safety

WARNING
WARNING
Power Supplies for additional equipment MUST comply with IEC 60601-1.
!! WARNING:
DO NOT attempt to use different peripherals and accessories (brand and model;
WARNING
WARNING
connected via USB ports) other than approved and provided by GE Healthcare!
!! WARNING:
The ultrasound system is an extremely sensitive and complex medical system.
Any unauthorized peripherals may cause system failure or damage!
The Voluson E8 / Voluson E6 is equipped with an isolation transformer to provide the required
separation from mains for both, the system and the auxiliary devices.
One AUX main outlet is located at the primary power supply. It is used for connecting the threefold/
fourfold splitter whose outlets are led to the shelves intend for auxiliary devices (e.g., printers).
BT Version: BT09 only: An AUX main outlet is accessible on the back of the control console.
The IEC 60601-1-1 standard provides a guideline for safely interconnecting medical devices in systems.
"Equipment connected to the analog or digital interface must comply with the respective IEC/UL
standards (e.g. IEC 60950 / UL 60950 for data processing equipment and IEC 60601-1 / UL 60601-1
for medical equipment).