Minimum Configuration to Boot/Scan

7-7-1
Minimum Configuration to Scan
1.) Connect the minimum configuration of cables to scan:
a.) Console
b.) USB UI (User Interface) and USB Hub
c.) Voluson E8: DVI Cable (Digital Visual Interface) from Graphic Card DVI Out to RTV DVI In
d.) Monitor
e.) Kontron Dual-Core BT10 only: RTV
BT Version:
Please observe that cable-minimum configuration depend on BT version and the currently installed
BT-Version:
PC-Motherboard version of the Voluson E8 / Voluson E6 system.
2.) Connect the Mains Power cable and mount the pull-out protection (see:
3.) Connect the Power Cable to an appropriate mains power outlet and switch ON the Circuit Breaker.
4.) Connect a probe.
5.) Press the ON/OFF Standby button on the Control Panel to boot up the System.
6.) Start an User Program.
KONTRON
Dual-core
version
e.g., Voluson E8
7
DRAFT (A
EVISION
Voluson E6: DVI Cable (Digital Visual Interface) from ADD-On DVI Out to RTV DVI In
Figure 7-22 cable- minimum configuration (BT09)
Chapter 7 - Diagnostics/Troubleshooting
23, 2012)
UGUST
a
b
c
d
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
Figure 4-4 on page
4-4).
a
DFI Dual-core
version
e.g., Voluson E6
b
c
d
7-19
E6