Table 10-16 Testing Power Conditions
Table 10-17
Measurements should be recorded from each lead and lead combination under each set of test
conditions specified in:
Table 10-10 on page
Table 10-11 on page
Record all data on the Inspection Certificate.
Table 10-17 Typical Data Format for Recording Patient Lead to Ground Leakage
System Under Test:
System Power
Mains on
RF on
NOTE:
Values in italics font are given as examples only.
10-20
7
DRAFT (A
EVISION
ECG Power
ON
ON
ON
OFF
OFF
OFF
OFF
shows a typical format for recording the patient lead to ground leakage current.
10-13, or:
10-13, as applicable.
________________________
Test Conditions
Grounding/PE
open
open
Section 10-7 - Electrical Safety Tests
23, 2012)
UGUST
Meter's Polarity Switch
NORM
REVERSE
REVERSE
NORM
NORM
REVERSE
REVERSE
Patient Lead or Combination Measured
Right Arm
Left Arm
(RA)
(LA)
V
OLUSON
Meter's Neutral Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
Date of Test:
_____________________
Left Leg
(LL)
E8 / V
E6
OLUSON
S
M
ERVICE
ANUAL
RA+LA+LL