OP Panel TGC Assy

Purpose: This is a description on how to remove and replace OP Panel TGC Assy.
8-4-6-1
Tools
Common pillips screwdrivers.
8-4-6-2
Needed Manpower
1persons, 5 minutes + travel
8-4-6-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
Keyboard Assy should be removed before proceed. For more information, refer to KeyBoard Assy
on section 8-4-1 on page 15.
8-4-6-4
Removal Procedure
1.) Remove eight (8) TGC knobs from the upper side of the Keyboard. Refer to Figure 8-30.
2.) Disconnect the TGC connector from the back side of the keyboard.
3.) Unscrew four (4) screws and Remove the TGC Assy from the keyboard.
8-26
2
Figure 8-30 Removing TGC Knobs
1
2
Figure 8-31 Unscrew 4 screws and remove TGC Assy