IRECTION
Table 10-14 Chassis Leakage Current Test Condition
TEST
2
3
10-7-5-3
Data Sheet for Enclosure/Chassis Leakage Current
Table 10-15
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by system.
Record all data in the Electrical safety tests log.
Table 10-15 Typical Data Format for Recording Chassis Source Leakage
System Under Test:
System Power
off
off
on
on
10-18
7
DRAFT (A
EVISION
CONDITION
Monitor housing
(mounting screw for LCD Monitor housing,
rear of monitor)
Probe connector
(probe mounting receptacle on front of system)
shows a typical format for recording the enclosure/chassis leakage current.
________________________
Test Conditions
Grounding/PE
closed
open
closed
open
Section 10-7 - Electrical Safety Tests
23, 2012)
UGUST
Date of Test:
Measurement/Test Point Location
Potential
Monitor
equilibrium
housing
connector
V
E8 / V
OLUSON
OLUSON
S
ERVICE
TEST POINT (image)
_____________________
Probe
connector
E6
M
ANUAL