GE Voluson E8 Service Manual page 270

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
6-5-2-1
Moving down (lower) the console - without booting up the system
If it is impossible to boot up the system, the user interface can be moved downwards (lowered) by
pressing 3 keys on the control console.
1.) Connect the Main Power Cable to the back of the system.
2.) Connect the Main Power Cable to a hospital grade power outlet with the proper rated voltage.
3.) Press 3 keys (see:
NOTE:
Shipping the Voluson E8 / Voluson E6 ultrasound system in its original packaging is only possible when
the system is lowered to its minimum height with monitor flapped and locked down, and when the control
console is centered and locked in position (see Figure on
6-5-2-2
Moving up (lift) the console - without booting up the system
BT Version: BT13 (Whanam console) only:
If it is impossible to boot up the system, the user interface can be moved upwards (lifted) by pressing
4 keys on the control console
1.) Connect the Main Power Cable to the back of the system.
2.) Connect the Main Power Cable to a hospital grade power outlet with the proper rated voltage.
3.) Press 3 keys + BF (see:
upwards.
6-8
7
DRAFT (A
EVISION
Figure 6-8
below) on the control console simultaneously to move it downwards.
Figure 6-8 keys to lower/lift the control console
Figure 6-8
Section 6-5 - Control Console Positioning
23, 2012)
UGUST
page
3-6).
above) on the control console simultaneously to move it
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
E6