GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
4-5-1
Save Image Settings Only / Save Small Backup (Scan Settings)
The Image/Scan Settings contain:
•
Application Settings
•
User Programs
•
Auto Text
•
3D/4D Programs
•
Scan Assistant
•
Measure Configuration
•
Biopsy Lines
1.) Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2.) Press the UTILITIES key on the control panel.
3.) In the "Utilities" menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
4.) On the left side of the screen select BACKUP and then click the
5.) Click the SAVE button of the "Image Settings Only" resp. "Small Backup (Scan Settings)" group.
6.) Choose the media and click the SAVE button.
7.) Select the NEW FILE... key and enter a file name (without extension).
8.) Click the OK key to start the process. When the saving has been completed, click OK.
7
DRAFT (A
EVISION
Figure 4-23 Save window
Chapter 4 - Functional Checks
23, 2012)
UGUST
V
E8 / V
OLUSON
OLUSON
S
ERVICE
SYSTEM CONFIGURATION
tab.
Choose media
and
then click [Save]
E6
M
ANUAL
4-33