Replacement of the Probe Holder for Endocavity probes

8-11-1
Manpower
One person, 15 minutes
8-11-2
Tools
Phillips screwdriver PH1 and PH2
8-11-3
Probe Holder (endocavity) - Removal Procedure
1.) Loosen the 4 screws and then remove the probe holder for endocavity probes.
8-11-4
Probe Holder (endocavity) - Installation Procedure
1.) Place the new probe holder at its original position and fasten it with the 4 screws.
Section 8-12

Replacement of the Trackball Ring

8-12-1
Manpower
One person, 5 min.
8-12-2
Trackball Ring - Replacement Procedure
1.) Remove the fixation ring by turning it counterclockwise.
rotate clockwise to
mount the fixing ring
2.) Mount the fixation ring by turning it clockwise.