Replacement Of The Caps For Hardkeys; Replacement Of Circle Key Caps Only - GE Voluson E8 Service Manual

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
Section 8-15

Replacement of the Caps for Hardkeys

NOTE:
The FRU part depends on Console (RTU).
Console version (Rafi or Whanam) is shown in the System Info page (see:
NOTICE
Replacement procedure depends on key caps that have to be replaced!
• if just the circle key caps have to be replaced, refer to:
!! NOTICE:
• if trackball buttons or mode key slices have to be replaced too, please contact your local distributor
or GE service representative.
8-15-1

Replacement of Circle Key Caps only

NOTE:
The FRU part depends on Console (RTU).
Console version (Rafi or Whanam) is shown in the System Info page (see:
8-15-1-1
Manpower
One person, approx. 1 minute/cap
8-15-1-2
Tools
small-sized slotted screwdriver or tweezers
8-15-1-3
Circle Key Caps - Replacement Procedure
1.) By means of a small slotted screwdriver, carefully push against the circle hardkey cap.
2.) Lift the cap, until it is completely loosened from its base.
3.) Place the new hardkey cap down until it snaps into position.
7
DRAFT (A
EVISION
Figure 8-25 push against the circle cap and lift it
Chapter 8 - Replacement Procedures