How to use the Auto Tester program
The following screen will appear.
"DISC" (Save)
icon
10.)Select the "Disc-Save" icon. The file will be saved in D:\Export\AutotesterScripts\*.*.
11.)Press ALT + D to export system data and the needed logs.
Operation see:
The standalone recorded "Auto Tester" file makes only sense with the ALT + D log!
NOTICE
To analyze the workflow the exact date and time of occurrence needs to be documented!
12.)If desired, save the script file also on DVD/CD. Therefore:
-
-
-
13.)Close the "Auto Tester" program.
7-6-1
Limitation of the Auto Tester
The following information will not be recorded. Depending on the moment the Auto Tester is activated,
this must be provided by the customer or the field engineer.
Which probe is in use and which probes are connected?
Which Mode is activated?
Which peripherals are connected (Dicom, Printer, etc..)?
Collected information from all steps above + exported system data and needed logs (ALT + D) can be
sent to the Make Center. With this information the Make Center can see how the customer is using the
system and reproduce potential failures.
NOTE:
Providing all information at once will help the Make Center find the root cause and speeds up finding a
solution for the customer.
7-18
7
DRAFT (A
EVISION
Figure 7-21 Auto Tester Finished
Section 7-4-2-1 "Export System Data (by pressing the ALT + D key)" on page
Insert an empty DVD/CD+R/RW in the Drive and select the "Burn on CD" icon.
Enter a Filename.
After the DVD/CD write is finished click the OK button.
Section 7-6 - How to use the Auto Tester program
23, 2012)
UGUST
(cont'd)
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
"DVD/CD" (Burn)
icon
7-10.