Basic Probe Cleaning and/or Disinfection (cont'd)
CAUTION
CAUTION:
Please be aware of the
sensitive probe head.
TAKE EXTREME CARE!
10-5-5-4
Disinfection by means of the RIC-Holder
Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when
transporting the system with the probe attached, or during the disinfection process.
Inadequate handling may lead to dead elements, shocked head mechanics, etc.
The RIC-Holder (especially developed for RIC Real-time 4D endocavity probes) guarantees that the
sensitive probe head does not hit the bottom of the container during the disinfection procedure.
NOTE:
Instructions are supplied with each RIC-Holder (KTZ225469).
OK
OK
RIC-Holder
7
DRAFT (A
EVISION
NEVER place or store a
probe on its scan head!
RIC-Holder
Chapter 10 - Care & Maintenance
23, 2012)
UGUST
OK
OK
V
E8 / V
OLUSON
OLUSON
S
ERVICE
When
disinfecting a probe, ensure
that there is sufficient space
between the probe and the
Endocavity soaking cup kit
RIC-Holder