During A Normal Boot, You May Observe - GE Voluson E8 Service Manual

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
3-6-2-3

During a normal boot, you may observe

A.) Power is distributed to Peripherals, Control panel, Monitor, Front-End and
Back-End Processor.
B.) The Back-End Processor and rest of the scanner starts with the sequence listed in following steps:
1.) First of all, the BIOS version is shown on the monitor.
2.) Afterwards the "Boot Screen" is displayed. (Voluson is highlighted,
3.) Back-End Processor is turned ON and starts to load the software.
4.) The Start Screen (Voluson) is displayed on the monitor.
5.) Start-up progress bars indicating software loading procedures, are displayed on the monitor,
6.) The software initiates and sets up the Front-End electronics and the rest of the scanner
7.) The Keyboard backlight is lit.
8.) As soon as the software has been loaded, the 2D screen is displayed on the monitor.
7
DRAFT (A
EVISION
Figure 3-27 Boot screen (e.g., BT12)
as shown in
Figure 3-28
Figure 3-28 GE Healthcare wallpaper with progress bar
(incl. the clicking sound of the relays on the RTF board).
Chapter 3 - Setup Instructions
23, 2012)
UGUST
below.
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
Figure 3-27
below).
3-33
E6