Preparation And Notes For Software Upgrade Procedure - GE Venue Fit Technical Publication

Table of Contents

Advertisement

P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
3.) Write down the following settings as it will be necessary to restore them at the end of the process:
-
-
-
-
8-3-2

Preparation and Notes for Software Upgrade Procedure

REMOVE ALL EXTERNAL DEVICES SUCH AS PRINTERS AND USB CONNECTIONS
WARNING
WARNING
BEFORE STARTING THE UPDATE PROCEDURE.
When updating the system software from a previous version, it is necessary to contact your local
OTR department to receive an appropriate software password. Make sure you specify the system
serial number located on the Venue Fit™ unit.
It is recommended to backup the data and configuration before starting the update procedure.
At the site, perform the following steps before you start the upgrade procedure:
1.) Perform Disk Management for all database records and select the "Copy" option
(for instructions, refer to the Venue Fit™ User Manual on your Document Media).
NOTE:
It is recommended that Disk Management be executed using the "move" option to a network
repository.
2.) Perform a full back-up for Patient Archive and System Configuration
(for instructions, refer to the Venue Fit™ User Manual on your Document Media).
3.) Write down the following settings as it will be necessary to restore them at the end of the process:
-
-
-
-
8-40
TCP/IP address
Network printer (if present)
Wireless network settings (if present)
InSite ExC settings
TCP/IP address
Network printer (if present)
Wireless network settings (if present)
InSite settings
Chapter 8 - Replacement Procedures
VENUE FIT™ SERVICE MANUAL

Advertisement

Table of Contents
loading