Solar 7000 Color Monitor
Color Display
Calibration
Revision J
CALIBRATION: Solar 7000 Color Monitor
The color CRT display of the Solar 7000 monitor is accurately adjusted at
the factory prior to shipment. If the CRT, CRT controller PCB, or
deflection PCB is replaced, it will be necessary to calibrate the display.
The following steps describe a simple method to calibrate the color
display using special displays generated by the monitor in the SERVICE
MODE menu.
To perform this procedure, you will need the special tools and items listed
below. Equivalent equipment may be substituted.
Required Tools/Special Equipment
Item
Very small Blade Screwdriver
Small Hex Coil Tuning Tool
Multimeter (DMM)
WARNING
To avoid electrical shock, do not touch the large heatsink
connected to the deflection PCB, pn 404642-002, while
the unit is ON.
Due to high voltage on the deflection PCB, use an
insulated screw driver at all times while making
adjustments.
WARNING
Make sure the monitor is not monitoring a patient.
Patient monitoring will be interrupted.
Solar 7000/8000/View Patient Monitor
414993-001
Manufacturer and
Part Number/Model
–
–
Fluke 8060A
5-3