Monitoring
Removing the flow sensor
1 Remove the tube connector.
2 Put on a sterile glove.
3 Seal the flow sensor. This ensures that the
requirement for calibration (flow = 0) is met.
Performing calibration
Press the Start key.
The calibration is completed after approx.
1 second. A message is displayed on the screen.
After calibration
Connect the tube connector.
If calibration was not successful
Replace the flow sensor insert, see page 42.
Replace the flow sensor cable.
96
Deactivating and activating flow
monitoring
WARNING
Risk of patient injury
The device cannot adequately monitor the
patient if flow monitoring and volume
monitoring are deactivated.
Provide appropriate substitute monitoring
immediately.
WARNING
Risk of patient injury
No apnea monitoring takes place when flow
monitoring is deactivated.
Use independent apnea monitoring.
CAUTION
Restricted ventilation functions
Patient-triggered ventilation is not possible if flow
monitoring is deactivated.
Reactivate flow monitoring as soon as possible.
Instructions for use Babylog 8000 plus SW 5.n