Download Print this page

Dräger Babylog 8000 plus Instructions For Use Manual page 30

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Operating concept
Exceeding setting limits
If the setting limit of a ventilation parameter is
exceeded, the LED for that setting knob flashes
and a message is displayed on the screen.
 Acknowledge the message by pressing the OK
key and set the ventilation parameter with the
setting knob.
Setting PEEP < 2.5 mbar (2.5 cmH
 Pull on the PEEP/CPAP setting knob and
simultaneously turn it to the left to the range
marked in red. Set the value for PEEP.
30
Keys with fixed and variable functions
G
F
O)
2
E
A
B OK
C Vent.
Mode
D Vent.
Option
E 6 variable
keys
F Cal. Con-
fig.
G man. Insp. Triggers manual inspiration, see
Setting the ventilation mode
1 Press the Vent. Mode key (C).
2 Press the key (E) for the desired ventilation
mode, e.g., SIMV.
3 Press the On key (E).
The ventilation mode is active.
Instructions for use Babylog 8000 plus SW 5.n
Suppresses the acoustic alarm
signal for 2 minutes
Acknowledges messages and
alarm messages
Opens the ventilation mode dia-
log window
Opens the additional settings dia-
log window
Selects functions or settings,
changes ventilation modes
Opens the calibration and config-
uration dialog window
page 69
A
B
C
D

Advertisement

loading