Dräger Babylog 8000 plus Instructions For Use Manual page 12

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

For your safety and that of your patients
Product-specific safety information
WARNING
Risk of incorrect use
This medical device is intended to be used
only by the target group "users".
WARNING
Risk of suffocation following device failure
If the gas inlet for the emergency breathing
valve is obstructed, the patient will not be
able to breathe spontaneously in the case of a
device failure.
Do not obstruct the gas inlet for the
emergency breathing valve.
WARNING
Risk of malfunction
Unallowed modifications to the medical
device lead to malfunctions.
This medical device must not be altered
without permission from Dräger.
WARNING
Risk of patient injury
During magnetic resonance imaging the
correct functioning of the medical device can
be impaired.
Do not use the medical device during
magnetic resonance imaging.
WARNING
Risk of patient injury
In hyperbaric chambers the correct
functioning of the medical device can be
impaired.
Do not use the medical device in hyperbaric
chambers.
12
WARNING
Risk of patient injury with PEEP <2.5 mbar
(2.5 cmH
O)
2
If the PEEP is set to <2.5 mbar (2.5 cmH
and the following settings are also made, the
integrated pressure monitoring cannot detect
an alarm situation during disconnection or
extubation and their immediate
consequences:
– Pinsp <10 mbar (10 cmH
– Ventilation mode with volume guarantee
– Change from a ventilation mode with VG
into the ventilation mode IPPV, without
switching off VG
From a PEEP of 2.5 mbar (2.5 cmH
of an undetected alarm situation increases
with decreasing PEEP values.
Use external ventilation monitoring as well as
the following types of monitoring with narrow
alarm limits:
– SpO
monitoring
2
– Bradycardia monitoring
– TcO
/TcCO
monitoring
2
2
WARNING
Risk of eye injury
With neonates, the administration of
increased O
concentrations can lead to
2
retinopathy of prematurity.
Use additional monitoring, e.g., external SpO
monitoring.
WARNING
Risk of incorrect NO dosing
If a device for nitric oxide (NO) delivery
without internal NO monitoring is used,
patient monitoring is not guaranteed.
Monitor the NO concentration separately.
Instructions for use Babylog 8000 plus SW 5.n
O)
2
O)
2
O) the risk
2
2