Signal output at the impulse output
The signal
at the impulse output displays
events during ventilation, e.g., every mandatory
breath or every triggered mandatory breath. The
voltage can only have high levels (H) and low
levels (L).
Depending on the setting, the pulse takes the
following path:
– Mandatory breath:
H during the mandatory breath, otherwise L.
– Triggered mandatory breath:
H during a triggered mandatory breath,
otherwise L.
– Alarm:
L during an alarm condition, otherwise H.
Analog 1
Analog 2
1 Connect the printer (input resistance ≥ 1 MΩ)
using the cable 8306487.
2 Select the signal, see page 103.
Instructions for use Babylog 8000 plus SW 5.n
Connecting a printer
Any of the following printers can be used::
– Epson LX 300
– Epson FX 870 with serial port
– HP laser printer with Epson emulation module
and serial port
Use other printers only after consulting Dräger.
1 Connect the printer using the cable 8306489.
2 Configure the RS232 interface, see page 104.
NOTE
The RS232 interfaces of the printer and
Babylog 8000 must be configured such that they
are compatible.
Operation
77