Dräger Babylog 8000 plus Instructions For Use Manual page 66

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Operation
Ventilation with separate expiratory flow
The additional setting VIVE can be used to set the
expiratory flow
in addition to the inspiratory
exp
flow
.
insp
1 Press the Vent. Option key.
2 Press the VIVE key.
3 Set the expiratory flow
keys.
4 Press the On key.
Ventilation with Volume Guarantee
CAUTION
Undetected worsening of the patient's condition
When using pressure-controlled ventilation, the
ventilation parameter VT is an important indicator
of the patient's condition, especially of changes in
lung compliance. Volume Guarantee reacts
automatically to changes in VT, in order to
maintain Vtset at the set value even in the event
of changing lung compliance. The inspiratory
pressure is adjusted automatically.
In order to detect changes in the patient's
condition, and especially lung compliance, early,
the user must not only take the VT into
consideration, but also the automatic adjustment
of the inspiratory pressure.
When Volume Guarantee is activated, the
inspiratory plateau pressure between Pinsp and
PEEP is automatically regulated so that the set
tidal volume Vtset is applied.
1 Set the pressure plateau, see page 64.
2 Press the Vent. Option key.
66
using the
or
ex
3 Press the VG key.
4 Set the tidal volume Vtset using the
keys.
5 Press the On key.
When the device is switched on and Volume
Guarantee is already activated, check the setting
for Vtset and adjust if necessary.
Deactivating Volume Guarantee
Volume Guarantee must be deactivated when the
device is switched to the IPPV ventilation mode.
Instructions for use Babylog 8000 plus SW 5.n