Printing waveforms
The following waveforms are printed graphically:
– Airway pressure
– Flow
– Tidal volume
1 Open the dialog window by pressing the Cal.
Config. > Print keys.
2 Press the Select key repeatedly until Graphics
is highlighted.
3 Start the print process by pressing the Start
key. The key function changes to Stop.
Canceling the print process:
 Press the Stop key.
Printing all data
The following data are printed:
– Report
– Trend
– Graphics
1 Open the dialog window by pressing the Cal.
Config. > Print keys.
2 Press the Select key repeatedly until All is
highlighted.
3 Start the print process by pressing the Start
key. The key function changes to Stop.
Canceling the print process:
 Press the Stop key.
Instructions for use Babylog 8000 plus SW 5.n
Transferring data to the patient monitor
Devices (e.g., monitor, PC), that work with the
BabyLink transfer protocol can be connected.
Additional information can be found in the
BabyLink instructions for use.
For information on configuring the interface, see
chapter ''Configuring the data interfaces'' on
page 101.
Performing data transfer
Condition: The monitor must be connected to
Babylog 8000 with the cable 8306488.
1 Open the dialog window by pressing the Cal.
Config. > Print keys.
2 Press the Select key repeatedly until
BabyLink is highlighted.
3 Start the data transfer by pressing the Start
key. The key function changes to Stop.
Canceling data transfer:
 Press the Stop key.
Operation
79