Dräger Babylog 8000 plus Instructions For Use Manual page 89

Intensive care ventilator for neonates
Hide thumbs Also See for Babylog 8000 plus:

Advertisement

Displaying measured volume values for high-
frequency ventilation (HFV)
Condition: High-frequency ventilation option must
be available.
1 On the main screen, press the Meas key.
2 Press the HFVol key.
The following measured values are displayed
(example):
MVim
Inspiratory measured minute volume
from mandatory ventilation [L/min]
DCO
Gas transport coefficient [mL
2
DCO
= VThf
2
f: Frequency of high-frequency pulses
[Hz]
VTim
Inspiratory measured tidal volume of
mandatory ventilation [mL]
VThf
Inspiratory measured tidal volume of
high-frequency ventilation [mL]
Instructions for use Babylog 8000 plus SW 5.n
2
/s]
2
x f
Displaying all set values and measured
values
Displaying all set values
 On the main screen, press the Values key.
The Set1 dialog window is displayed.
Displaying additional set values:
 Press the Set2 key.
Displaying all measured values
1 On the main screen, press the Values key.
2 Press the Meas1 key.
Displaying additional measured values:
 Press the Meas2 key.
Trends
89