Replacing the flow sensor insert
The insert of the flow sensor must be replaced
when the following alarm message is displayed:
Flow measurement disturbed
Measurement switched off
1 Remove the plug (A) of the flow sensor cable
from the flow sensor.
2 Gently press the knobs (B) on both sides while
pulling the insert out of the flow sensor
housing.
3 Push the new insert (C) in until it engages.
Ensure that the two markings are aligned.
4 Connect the plug (A) of the flow sensor cable to
the flow sensor.
5 Calibrate the flow sensor, see page 95.
Instructions for use Babylog 8000 plus SW 5.n
A
C
Connecting the test lung
The test lung consists of the following
components:
– Bellows:
Compliance: 0.5 mL/mbar (0.5 mL/cmH
– Endotracheal tube CH 12:
Length: approx. 165 mm (6.5 in)
– Connector
B
Connecting the test lung
 Insert the test lung in the patient connection
port of the breathing circuit.
Assembly and preparation
O)
2
47