Technical data
Recommended safety clearance for portable and mobile high-frequency
communication equipment
The safety clearances listed in the following
comply with IEC 60601-1-2.
Max. P
150 kHz to 2.5 GHz
EIRP
(watts)
0.03
0.32 m (1.1 ft)
0.10
0.58 m (1.9 ft)
0.17
0.76 m (2.5 ft)
0.20
0.82 m (2.7 ft)
0.25
0.92 m (3.0 ft)
0.41
1.2 m (3.9 ft)
0.82
1.7 m (5.6 ft)
1.00
1.8 m (5.9 ft)
1.64
2.4 m (7.9 ft)
3.3
3.3 m (11 ft)
Reduced separation distances to portable and mobile radio frequency
communication devices
The following separation distances are based on
additional tests performed by Dräger to determine
the minimum separation distances absolutely
necessary. These reduced separation distances
are valid only for mobile radio frequency
communication devices using the standards listed.
Mobile radio frequency communication device using ...
GSM 850, GSM 900, RFID 868 MHz (limited to 2 W ERP)
GSM 1800, GSM 1900 (limited to 1 W ERP)
UMTS, DECT (limited to 0.25 W ERP)
Bluetooth, WLAN 2450, RFID 2450 (limited to 0.1 W ERP)
150
All other frequen-
Examples
cies
0.96 m (3.2 ft)
WLAN 5250 / 5775 (Europe)
1.8 m (5.9 ft)
WLAN 2440 (Europe)
2.3 m (7.6 ft)
Bluetooth, RFID 2.5 GHz
2.5 m (8.2 ft)
WLAN 5250 (not in Europe)
2.8 m (9.2 ft)
UMTS mobiles
3.5 m (12 ft)
Cordless DECT devices
5.0 m (16 ft)
RFID 13.56 MHz
5.5 m (18 ft)
WLAN 5600 (not in Europe)
7.1 m (23 ft)
GSM 1800 / GSM 1900
10 m (33 ft)
GSM 900 mobile phones, RFID 868 MHz
Instructions for use Babylog 8000 plus SW 5.n
Separation distance
0.67 m (2.20 ft)
0.38 m (1.25 ft)
0.19 m (0.62 ft)
0.07 m (0.23 ft)