Reprocessing list
Applicable to non-infectious patients.
The reprocessing list contains approximate values
only. The instructions of the hospital's infection
control officer responsible have priority.
Non-critical medical devices
Items which can be repro-
cessed
Ventilator
Trolley
Hinged arm
Flow sensor cable
Compressed gas hoses
Semi-critical medical devices
Items which can
Recom-
be reprocessed
mended repro-
cessing
intervals
Breathing hoses
Per
patient/weekly
Y-piece
Per
patient/weekly
1)
Water traps
Water trap con-
tainer
Expiratory valve
Per
patient/weekly
Muffler
Per
patient/weekly
Housing of the
Daily
ISO 15 flow sensor
Y-piece flow sen-
Daily
sor
Instructions for use Babylog 8000 plus SW 5.n
Recommended repro-
cessing intervals
Per patient
Per patient
Per patient
Per patient
Per patient
Preclean-
Machine
ing
cleaning
and disin-
Yes
Yes
Yes
2)
Yes
Yes
According to the corresponding Instructions for use and section
''Manual reprocessing of the Y-piece'' on page 118
Cleaning, disinfection and sterilization
Cleaning
Yes
Yes
Yes
Yes
Yes
Manual
Cleaning
fection
Yes
According to the corre-
sponding instructions
for use
Yes
Possible
Yes
Possible
Yes
Possible
Yes
Possible
Manual
Disinfection
Yes
Yes
Yes
Yes
Yes
Steriliza-
tion
Disinfec-
tion
Yes
Possible
Yes
Possible
Yes
Possible
Yes
Possible
Yes
117