Disconnecting the breathing hoses
CAUTION
Damage to the breathing hoses
When removing the breathing hoses, hold them
at the connection sleeve and not at the coil
reinforcement.
1 Remove the breathing hoses from the
inspiratory port and the expiratory port.
2 If in use: remove the water trap from the
breathing hose.
3 Remove the water trap container from the
water trap and empty it.
4 Remove the bacterial filter and dispose of it in
accordance with the corresponding instructions
for use.
Reprocessing the breathing circuit
 Reprocess the breathing hoses, the water trap,
and the water trap container in accordance with
the reprocessing list, see page 117.
Instructions for use Babylog 8000 plus SW 5.n
Cleaning, disinfection and sterilization
Removing the expiratory valve
A
1 Tilt the rocker lever (A) up.
2 Pull the expiratory valve forward to remove it.
3 Remove the muffler from the expiratory valve.
Reprocessing the expiratory valve
 Reprocess the expiratory valve and muffler in
accordance with the reprocessing list, see
page 117.
Dismantling the accessories
 Dismantle and reprocess the breathing gas
humidifier and the Aeroneb Pro nebulizer in
accordance with the corresponding instructions
for use.
 Dismantle and dispose of the bacterial filter in
accordance with the corresponding instructions
for use.
113