Waveform display (on screen)
Range
Resolution
Flow and volume measurement
Sensor
Neonatal flow sensor Y-piece (8410185)
Dead space
Resistance
Neonatal flow sensor ISO 15 (8411130)
Dead space
Resistance
Flow
Range
Tidal volume
Range
Resolution
Accuracy
Minute volume
Range
Resolution
Accuracy
(Sensor in standard breathing circuit,
endotracheal tube >3 mm (0.12 in))
Leakage
Range
Resolution
Spontaneous portion of minute volume
Range
Resolution
Instructions for use Babylog 8000 plus SW 5.n
–10 to 100 mbar (–10 to 100 cmH
–2.5 to 25 mbar (–2.5 to 25 cmH
(0.5 cmH
O)
2
–5 to 50 mbar (–5 to 50 cmH
(1 cmH
O)
2
–10 to 100 mbar (–10 to 100 cmH
(2 cmH
O)
2
Hot wire anemometer
1.7 mL
≤12 mbar (≤12 cmH
O) at 30 L/min
2
0.9 mL
≤11 mbar (≤11 cmH
O) at 30 L/min
2
0.2 to 30 L/min
VT
0 to 999 mL
0.1 mL
≤5 mL: ±0.5 mL
>5 mL: ±10 % of measured value
MV
0 to 30 L/min
0 to 0.99 L/min: 0.01 L/min
1 to 9.9 L/min: 0.1 L/min
10 to 30 L/min: 1 L/min
≤2 L/min: ±0.2 L/min (NTPD)
>2 to 30 L/min: ±10 % (NTPD)
Leak
0 to 100 %
1 %
spont
0 to 100 %
1 %
Technical data
O)
2
O): 0.5 mbar
2
O): 1 mbar
2
O): 2 mbar
2
135