Principles of operation
Monitoring of high-frequency ventilation
As in conventional ventilation, the waveforms for
pressure and flow are measured and can be
displayed. The following are measured specifically
for high-frequency ventilation:
2
DCO
= VThf
x f
2
Transport coefficient for CO
, which, analogous to
2
the minute volume in conventional ventilation, is a
measure of high-frequency ventilation.
VThf
Tidal volume of the high-frequency pulses,
averaged over multiple cycles
MVim
Minute volume applied by IMV breaths, measured
on the inspiratory side
VTim
Tidal volume applied by IMV breaths, measured on
the inspiratory side
164
Instructions for use Babylog 8000 plus SW 5.n