Maintenance
Inspection
Perform inspections at regular intervals and
observe the following specifications.
Checks
Inspection and safety checks
1) Designation applies to the Federal Republic of Germany; corresponds to the "Recurring safety inspection" in the Republic
of Austria
Safety checks
The safety checks are no substitute for the
preventive maintenance measures (including
preventive replacement of wear parts) indicated by
the manufacturer.
CAUTION
Risk of medical device failure
If safety checks are not performed on a regular
basis, the proper operation of the medical device
can be compromised.
Perform safety checks at the indicated intervals.
1 Check accompanying documents:
– Instructions for use are available
2 Check that the device combination is in good
condition:
– All labels are complete and legible
– No visible damage to the following parts:
– Trolley and brakes
– Housing parts
– Gas inlets
– Hoses and cables
– Fuses which are accessible from the
outside are in compliance with the specified
values
– Country-specific labeling of gas types
3 Using the instructions for use, check that all
components and accessories needed to use
the product are available.
122
Interval
1)
Every 6 months
Personnel responsible
Service personnel
4 Check safety features:
– Correct functioning of the emergency
breathing valve:
Before the check, switch off the device.
Pressure: max. –4 mbar (or hPa or cmH
Instructions for use Babylog 8000 plus SW 5.n
O)
2