Device outputs
Analog and digital interface (optional)
Analog-1 and Analog-2 outputs
Signal delay
Output resistance
Plug connection
Impulse output
High-level voltage
Low-level voltage
Output resistance
Plug connection
RS232 interface
Printer connection
Pin configuration
Monitor connection
Pin configuration
Instructions for use Babylog 8000 plus SW 5.n
All outputs on the analog/digital interface are electri-
cally isolated from the device electronics. Electric
strength: 1.5 kV.
The outputs are short circuit proof. Each voltage is
output via a 12-bit DA transducer with downstream
low-pass filter.
Electronic filter circuits in the device delay the sig-
nals for airway pressure and flow by approx. 15 ms
compared to sensor signals. When a separate mea-
suring device is used and its signal is compared to
the analog output of the device, this delay must be
taken into consideration.
Approx. 10 kΩ
SMB-Subclic
The output is short circuit proof.
5 V ±0.5 V, no load
0 V ±0.5 V, no load
<5 kΩ
SMB-Subclic
Level in accordance with DIN 66020
With 8306489 printer cable only
With 8306488 monitor cable only