Displaying waveforms and measured values
Displaying waveforms
Displaying the airway pressure waveform Paw
1 On the main screen, press the Graph key.
2 Press the Paw key.
A
B
A Pressure limitation Pinsp (dotted horizontal
line)
B Pressure axis scale (e.g., 50 mbar
(50 cmH
O))
2
C End of the set expiratory time TE (dotted verti-
cal line)
D Time axis scale (e.g., 2 s)
Displaying the flow waveform
1 On the main screen, press the Graph key.
2 Press the Flow key.
E F
E Flow axis scale (e.g., 10 L/min)
F Zero line
G End of the set expiratory time TE (dotted verti-
cal line)
H Time axis scale (e.g., 2 s)
Instructions for use Babylog 8000 plus SW 5.n
C
D
G H
Freezing waveforms
1 On the main screen, press the Graph key.
2 Press the Stop key.
The waveform is immediately frozen.
Return to displaying the current waveform:
 Press the Stop key.
Displaying measured values
Displaying the measured pressure values
1 On the main screen, press the Meas key.
2 Press the Paw key.
The following measured pressure values from the
previous respiratory cycle are displayed
(example):
Peak
Peak pressure
Mean
Mean pressure
PEEP
End-expiratory pressure
Trends
87