Operation
Terminating operation
Switching the ventilator off
Condition: No patient must be connected to the
device.
1 Move the protective flap (A) on the rear of the
device to the side.
2 Press the mains switch (B) as far as it will go
and then release it.
The device is switched off.
3 Unplug the mains plug.
4 Switch off the breathing gas humidifier and
remove the mains plug.
80
Removing the compressed gas hoses
CAUTION
Contamination of the central gas supply system
Supply gases may flow back into the central gas
supply system due to small leakages in the
A
device or the compressed gas hoses.
B
After terminating operation, disconnect the
compressed gas hoses from the central gas
supply system.
CAUTION
Risk of personal injury
If the probes in the wall terminal units of the
central gas supply system are connected, the
compressed gas hoses are under pressure and
may injure the user when unscrewing them from
the ventilator.
The compressed gas hoses must only be
unscrewed from the ventilator after the probes
have been removed from the wall terminal units.
 Disconnect the probes of the Air and O
compressed gas hoses from the wall terminal
units of the central gas supply system.
Instructions for use Babylog 8000 plus SW 5.n