Cleaning, disinfection and sterilization
Items which can
Recom-
be reprocessed
mended repro-
cessing
intervals
Flow sensor insert Daily
Breathing gas
Per
humidifier
patient/weekly
Bacterial filter
Medication nebu-
1)
lizer
Parts for adapting
Per
patient/weekly
1) Keep spring-loaded valves (water trap, pneumatic medication nebulizer) open during reprocessing.
2) Nebulization may lead to increased deposits making it necessary to exchange the parts more often.
Manual reprocessing of the Y-piece
A
B
C
Manual cleaning:
1 Immerse the Y-piece in the solution and agitate
it slightly so that the air can escape.
2 Before the contact time begins and after the
contact time has elapsed:
 Insert and remove a pipe cleaner (A) ten
times vertically into each of the two
connection openings of the Y-piece (B) and
then, at an angle, insert and remove the
pipe cleaner ten times into both corners of
the opening for the insert (C).
 Fit a syringe (D) containing 20 mL of the
solution to each opening of the Y-piece.
Inject the solution three times.
118
Preclean-
ing
and disin-
According to the corresponding Instructions for use
According to the corresponding instructions for use
According to the corresponding instructions for use
According to the corresponding instructions for use
Yes
D
Machine
Manual
cleaning
Cleaning
fection
Yes
Possible
Perform manual disinfection in the same manner.
Instructions for use Babylog 8000 plus SW 5.n
Steriliza-
tion
Disinfec-
tion
Possible
Yes