MEDIBUS protocol
MEDIBUS is a software protocol for data transfer
between Babylog 8000 and other medical devices
(e.g., patient monitors) or other devices (e.g.,
computers for data management systems).
Requirements for the combination of Babylog 8000
and an external device, see ''Device combinations''
on page 9.
CAUTION
Incorrectly transferred data
All transferred data is for informational purposes
only and must not be used as basis for diagnostic
or therapeutic decisions.
Check the displays on Babylog 8000. The
MEDIBUS interface is not intended for use with a
"distributed alarm system" according to
IEC 60601-1-8:2012.
Observe the following documents:
MEDIBUS for Dräger Pediatric
Devices
Dräger RS 232 MEDIBUS, Protocol
Definition
Instructions for use Babylog 8000 plus SW 5.n
Connecting the external device for
MEDIBUS
Conditions:
– The appropriate MEDIBUS cable must be
used.
– Only devices with safety extra-low voltage
(SELV) must be connected to the COM
connection (serial RS232 interface).
 Connect the external device to the COM port
9029205
(A).
9028258
Configuring the interface
For a description of the process, see chapter
''Configuring the RS232 interface'' on page 104.